5.2 Standard Gas Calibration
To perform a Standard Gas Calibration, using the supplied calibration cap connect the
appropriate gas cylinder to the device.
• Press the button to navigate to Calibration Mode, the calibration icon will be shown, .
• Press and hold the button for 5 seconds, 'CAL' and the fresh air calibration icon will be
displayed on the screen,
• Press the button once until the calibration icon is shown, .
• Press and hold the button for 3 seconds to initiate the Standard Calibration. When the
calibration begins, a countdown (60 seconds or more depending on the gas type) will
appear on the screen.
%
V
%
X
%
If calibration fails,
X
display. If this continues, please contact the sales
representative or Crowcon customer support.
%
.
%
VOL
V
%
VOL
Once completed,
seconds on the display. The device will then return to
Gas Measurement mode.
%
If calibration fails,
X
If this continues, please contact the authorised
reseller or the Crowcon customer support team.
%
icon will appear on the
X
icon will appear for several
V
icon will appear on the display.
X
13