5. Calibration
⚠
CAUTION
For O
units only. Initial calibration is performed on all devices prior to shipment. Once received,
2
calibration should be performed monthly (or quarterly) depending on frequency of use.
5.1 Fresh Air Calibration
⚠
CAUTION
Calibration should be performed in a fresh-air environment that is free from the influence of other
gases (calibration is assumed to be performed in an environment with a concentration of 20.9%
oxygen). It is also recommended that calibration should not be performed in a confined space.
To perform a Fresh Air Calibration press the button to navigate to Calibration Mode, the
calibration icon will be displayed on the screen
• Press and hold the button for 5 seconds, 'CAL' and the fresh air calibration icon will be
seen on the screen
• Press and hold the button for 3 seconds to initiate the Fresh Air Calibration.
When the calibration begins, a countdown (starting at 10) will be displayed on the screen.
%
VOL
V
12
.
Press ESC to show Gas Measuremnt mode.
Once completed,
%
%
Fresh Air Calibration
Standard Gas Calibration
.
icon will appear on the LCD.
V
Clip SGD
Need help?
Do you have a question about the Clip O2 and is the answer not in the manual?