Clip SGD
Figure 2
Docking Info
• Serial S/N: Factory serial number
• Version: Docking Station firmware version
• Auto Power-Off : Docking Station automatic power-off time (Max 3600 seconds)
• Maximum Bump Time: Bump test time (seconds) – Default: 30 / Min: 20 / Max: 120
• Location: Configure the installed location
• Gas Expiration Date: Enter the expiration date of the gas cylinder.
• Gas Bottle Lot #: Enter the lot number of the gas cylinder.
• Info Read/Write: Retrieve or write settings from Docking Station
Enable/Disable Function
Check box to enable or disable functions; a tick means enabled.
• Calibrate: enables/disables Docking Station 'Calibration' button.
• Bump test: enables/disables Docking Station 'Bump Test' button.
• Update SGD Configuration: enables/disables Clip SGD configuration updates during
bump test or calibration.
• Update SGD Firmware: enables/disables Clip SGD firmware update during bump test or
calibration.
Docking Firmware Upgrade
• Allows Docking Station firmware to be upgraded – click 'Browse' to navigate to file.
• Click 'Write' to download file.