• Press and hold the button for 3 seconds, and the "tst" message will be displayed for 45
seconds (to cancel, press the button for one second).
• Apply the gas using the calibration cap and gas tube within the 45 second period. If
a Bump Test is not undertaken within this period the "bts" message will flash until a
successful test is conducted.
After the selected gas is applied and if the test is successful, the "SUC" message with
the
icon will be displayed for 3 seconds then immediately return to Gas Measurement
V
Mode. The calibration cap and gas tube can then be removed.
If the test fails, the "FA" message with
flash until a successful test is conducted.