Agilent Technologies 1260 Infinity VL User Manual page 162

Quaternary pump
Table of Contents

Advertisement

11
Identifying Cables
Remote Cables
Agilent Module to General Purpose
p/n 01046-60201
162
Wire Color
Pin Agilent
module
White
1
Brown
2
Gray
3
Blue
4
Pink
5
Yellow
6
Red
7
Green
8
Black
9
1260 Infinity QuatPump VL User Manual
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low