Electrical - Titan PULSITE LINK Instruction Manual

Hide thumbs Also See for PULSITE LINK:
Table of Contents

Advertisement

3 Electrical

The Pulsite
Link can be programmed using the USB connection via a computer.
®
Electrical and instrument connections are wired internally via 2 x PG7 Cable glands and 5mm
pitch terminal block connectors.
To access these connectors, simply undo the 4 rear screws and carefully remove the front
cover.
Description
Specification
15-24Vdc Regulated
Power Supply
0.1 amp
Pulse (max. 5kHz)
Flowmeter Input
Reed Switch (max. 400Hz)
Sensor Supply 5Vdc 50mA max room
temperature. (25mA at high temperature)
NPN/PNP 2.5kHz
Output Pulse
24V@20mA max.
Scalable pulse per litre
0-5Vdc / 0-10Vdc
Output Analog
16 Bit
Adjustable flow range
4-20mA
16 Bit
Adjustable flow range
© Titan Enterprises Ltd 2023
®
ACTIVE
www.flowmeters.co.uk
Terminal Connection
Power Input +V
+24VDC →
Power Input 0V
0VDC →
+VDC Supply to meter →
0VDC Supply to Flowmeter →
Pulse from flowmeter →
Flowmeter
Pull up/down resistor required for NPN/PNP –
see wiring diagrams
Pulse Output to PLC/display →
NPN or PNP
Pull up/down resistor required for NPN/PNP –
see wiring diagrams
Analog Output V Out
+VDC to PLC/Display →
Analog Output 0V
0VDC of circuit →
Analog Output mA Out
+mA to PLC/Display →
Analog Output 0V
0VDC of circuit →
Pulsite®Link_Instruction_Manual_0823
7
Flowmeter +5V Out
Flowmeter 0V
Digital Output

Advertisement

Table of Contents
loading

This manual is also suitable for:

370-008

Table of Contents