5.4 User Settings Tab

5.4.1 Calibration Settings

© Titan Enterprises Ltd 2023
®
Output Smoothing: This selects the rate that the analog
output and asynchronous pulse outputs are updated.
Chose Automatic and the Pulsite Link will adjust the
sampling response time to the fastest available from the Hz
pulse input.
Alternatively, the user can chose a single response time for
the flow range of operation. For continual reliable sampling
the slowest response time must be slower than the lowest
expected input rate.
Once the calibration data is entered the smoothing cutoff
will display the lowest flow that can be reliably measured.
Hovering your mouse over the
Settings tab will also show the Smoothing cut-off and
recommend you set above this value.
Setting the cut-off value below the recommended level
means that when the flow is between the set value and the
recommended value, the display will occasionally read
zero.
K Factor Values: Enter 1 to 10 sets of values of flow and
pulses per litre (K factor) as found in the calibration
certificate supplied with your flow meter or instrument
Units:
A drop down menu offers the choice of Litres, ml, gms,
kg, US gallon, Imperial gallon or Custom units i.e. blank
Note that the Custom Units setting will be equivalent to Litres if no
modifications are made to scale factor.
www.flowmeters.co.uk
Cut Off
value in the User
Pulsite®Link_Instruction_Manual_0823