Titan Interface Software opens but
shows not connected to device (no
spinning bars in bottom left of
screen)
Titan Interface Software is not
showing all Tabs (User Settings;
Flow Details; Advanced;
Datalogging)
ASCII not being seen/sent
© Titan Enterprises Ltd 2023
®
Two windows of the Interface
Software are open and
second window is masking
connected window
Damage to USB Connection
Software is placing the tab
window outside the monitor
area
ASCII Streaming not enabled
VPN not set up correctly on
receiving device
www.flowmeters.co.uk
Either drag window out of
the way or close second
window using task bar
Return to Titan Enterprises
for repair
Go to View on the top bar
menu and click Restore
Defaults. This restores the
Viewing defaults and re
docks all tabs to the main
window
Power Cycle the device
and/or connect Titan
Interface Software and exit
program before
disconnecting unit.
Set up VPN – See section 4
ASCII Data Stream via USB
Pulsite®Link_Instruction_Manual_0823