GE Innova IGS 620 Preinstallation Manual page 168

Table of Contents

Advertisement

Cable
R rectifier
5 x 10 mm²
L out /load
5 x 10 mm²
(1) Size and type may be adapted locally per local regulation.
NOTE:
Electrical contractor is responsible for providing the power cables from Hospital mains
to Large Display UPS and the ground cable from Hospital mains to Large Display
Cabinet as per local regulations.
168
Illustration 5-4: Power distribution - UL
CE UPS configuration
Recommended (1)
Type
3ph+N+PE
3ph+N+PE
Innova
IGS 620, Innova
TM
UL UPS configuration
Max gauge
Recommended (1)
AWG3
AWG6
AWG3
AWG6
IGS 630 Preinstallation Manual
TM
Direction 5694389-1-1EN, Revision 2
Type
3ph+N+PE
3ph+N+PE
1 Power Requirements
Max length
24 m (78ft
8in)
24 m (78ft