Injectors.
Heart monitoring equipment.
Crash cart.
2.3.4 Emergency Stop
It is recommended to install, in an accessible location, an additional EPO button that allows the
instantaneous switching off of all power from the System (EXCEPT THE INPUT CABLE OF
THE 3kVA UPS FOR CE CONFIGURATION WITHOUT FLUORO UPS), including UPS and
Fluoro UPS outputs.
To remove input power from THE INPUT CABLE OF THE 3kVA UPS (FOR CE
CONFIGURATION WITHOUT FLUORO UPS), turn OFF the PDB main breaker
Protect the Emergency Stop from accidental actuation.
2.3.5 Patient Environment Equipment
The components that may be installed within patient vicinity need to be medical equipment
("patient vicinity is defined in the standardization as a space within the room 1.83 m (6 ft)
beyond the perimeter of the examination table and extending vertically 2.29 m (7 ft, 6 in) above
the floor."). For the Innova System, the equipment are:
Table
C–arm
Monitors
Injector
2.3.6 Preference Cabinet locations
NOTE:
This is applicable for all types of cabinets (including PDB).
The clear width of the service area in front of the cabinet doors to insure electrical safety shall
be at least 0.9 m for all cabinets except the Large Display cabinet (1 meter is necessary for
Large Display cabinet). In cases where 2 cabinets are installed face to face (both sides of the
access way), the clear width shall be at least 1.2 m.
104
WARNING
OPTIONAL LARGE DISPLAY SECONDARY MONITOR MUST BE INSTALLED
OUTSIDE OF THE PATIENT VICINITY
Innova
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
TM
Direction 5694389-1-1EN, Revision 2