E331.E2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
4.1.2.
Installing the Outer Case
1.
Slide
the
outer
case
onto
the
monitor
chassis from
the rear of
the monitor,
so that the front of
the case is seated
in its channel
on the top and sides of the front
panel and the five mounting screw
holes in the rear of the case align
with
the
mounting
holes
in the
chassis.
Insert
five
Phillips-head
screws
with
washers
into the
mounting
holes,
and
tighten the screws
to
secure
the
outer
case
to
the
chassis.
NOTE:
Insure that all five screws
are
in place and tight, and that the outer
case is secured, to seal the unit from rf
or electromagnetic field interference.
4.2.
4.2.1.
1.
Power Supply, A140/A280
Refer to Figure 4-2.
Removing the Power Supply
Remove the monitor's outer case
following the procedure outlined in
Section 4.1.1.
Disconnect
the
power
cable
assembly
plugs
marked
X1, X2,
and X3 on the top edge
of the
power
supply
PC
board.
The
power supply is located between
the CRT shield and the left side
panel of the monitor chassis.
Remove and save the two Phillips-
head
screws
(A)
securing
the
power supply to the monitor side
panel.
Figure 4-2.
Location of Power Supply, A140/A280