ЕЗЗ1.Е2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
X
4.
REMOVING/INSTALLING SYSTEM
COMPONENTS
Introduction:
Before
attempting
to
remove the monitor cover or disassemble
the monitor always do the following for
personal safety and to protect equipment:
* Switch the main power OFF.
+ Disconnect
the
main
power
cable
and all other cables attached to the
monitor.
Ts
CAUTION: SIRECUST 1280/1281 monitors
contain electronic components
that can
be damaged by electrostatic discharge.
Open the monitor case only in a static-
protected environment. Observe standard
procedures for protecting the equipment
from static electricity.
NOTE:
Since
all
procedures
in this
O
Chapter
assume
that
only
the
part
or
subassembly
identified
in
the
major
Section
heading
is to
be
removed
or
redundancy
in the description
of some
procedures. "Remove" instructions assume
that
the
procedure
begins
with
a
completely
assembled
monitor
"Install"
instructions, assume that the monitor has
been disassembled to the extent that it is
ready to have the subassembly installed.
4.1.
Opening/Closing the Monitor
Refer to Figure 4-1.
4.1.1.
Removing the Outer Case
NOTE:
For personal safety and pro-
tection of equipment, observe all pre-
cautions
stated
in
the
Introduction
section of this Chapter.
1.
Remove and save the five Phillips-
head
screws
and
washers
securing
the
outer case
to the
monitor chassis.
2.
Slide the outer case towards the
rear
of the
chassis
and
lift it off
installed
(replaced),
there tends to be
the monitor.
ὧν
ITS
Figure 4-1.
Removing/installing Outer Case
Page 4-1