E331.E2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
Table 6-3.
Signal Summary (Continued)
A
SIGNAL
TEST POINT
BOARD OF ORIGIN
+15V EXTA
X16 pin 21 (ref. to pin 24): rear panel
Front End
+15V EXTB
X17 pin 21 (ref. to pin 24): rear panel
Front End
25K1
X16 pin 17 (ref. to pin 18): rear panel
Front End
25K2
X17 pin 17 (ref. to pin 18): rear panel
Front End
DCS A12
X3 pin 9 (ref. to pín 10): Front End board
SIREM
DCS B12
X3 pin 11 (ref. to pin 12): Front End board
SIREM
DCS C12
X4 pin 9 (ref. to pin 10): Front End board
SIREM
DCS D12
X4 pin 11 (ref. to pin 12): Front End board
SIREM
In addition,
also
consider
the following
information in Sections 6.4.1 and 6.4.2:
6.4.1.
Display Subsection Boards
If
the
troubleshooting
procedures
of
Section
6.7 indicate
a faulty Deflection
Board,
Z-Amp
Board,
or Power
Supply
Board, always verify that the Curve Board
output
signals
are functioning
properly
(see Section 6.8) before connecting the
Curve
Board
to
a
new
replacement
board(s).
A defective Curve
Board
can
damage other display subsection boards.
6.4.2.
Signal Processing and Control
Boards
If the
troubleshooting
procedures
of
Section
6.7
indicate
a
faulty
signal
processing and control board, check the
VME
connectors on the board for bent
pins and proper configuration of jumpers,
prior to PC board replacement. Refer to
Appendix C for jumper configurations.
6.5.
Signal Summary
Table 6-3 lists the test points and signals
used for troubleshooting a 1280 or 1281.
This signal summary is intended for quick
reference
purposes
only.
The
trouble-
shooting procedures described in Section
6.7 should be used as the primary source
of
information
for
troubleshooting
a
problem within a 1280 or 1281.
WARNING: | Hazardous
voltages
are
present while probing signals
listed in
Table
6-3.
Always
turn
power
OFF
whenever
connecting/disconnecting
PC
boards
and
cables,
checking
and
replacing fuses, or connecting and dis-
connecting equipment probes to or from
test points.
The high-voltage circuitry retains lethal voltage
for approximately 30 seconds after the monitor
has been switched OFF
6.6.
Troubleshooting Flowchart
The troubleshooting flow chart in Figure
6-1
on
pages
7a thru 7e,
illustrates a
process for isolating defective PC boards
or subassemblies. The chart is intended
as a quick reference only.
The troubleshooting procedures outlined
in Section
6.7 should
be used
as the
Page 6 - 5