F331 .E2501.061.21.02.02
System SIRECUST 1280/1281
Bedside Monitor - Service Manual
E-2.
Cardiac
Output
Test
Plug:
Construct
a cardiac
output test plug,
using
a
SIEMENS Art. No. 73 68 640 E2260 plug or equivalent. Refer to Figure E-3.
R
= 14 40.1% КО
Art. No. 55 59 331 B0512
R,
= 53.6 +0.1% КО
Art. No. 99 65 815 B0512
R,
= 291 +0.1% 0
Art. No. 99 65 831 B0512
R,
= 9.76 +0.1% kQ
Art. No. 99 65 856 B0512
R,
= 402 +0.1% Q
Art. No. 99 65 849 B0512
TEST PLUG
ンーーーーー
Figure E-3.
Cardiac Output Test Plug Assembly
1.
Connect and solder the resistor(s) as shown in Figure E-3 for the specified test
temperature.
2.
Install the heat shrink tubing insulation, and complete assembly of the simulator.
Page E-2