15
|
YES
KEY
COORDINATES
APPEAR
TO
DRIFT
OVER
TIME?
(6.71.4)
PERFORM
TOUCHSCREEN
ACCURACY
TEST.
REPLACE
TOUCHSCREEN
PROBLEM
WITH
AUDIBLE
TONES?
PROBLEM
PZ
WITH
SPEAKER
SOUNDS
(SPKR)
OR
PIEZO
SOUNDS
OPERATOR
YES
INTERFACE
LEDS
TOGGLE
FOR
KEY
REPLACE
68705
(J9)
GN
OPERATOR
INTERFACE
BOARD.
IF PROBLEM
IS
CHECK
SPEAKER
ASSEMBLY
AND
DISPLAY
PROCESSOR
BOARD
TO
ISOLATE
TO
BLOCK
K
CHECK
JUMPER
JP90
ON
MOTHER
BOARD.
CHECK
BATTERY/
MEMORY
BACKUP
BOARD
TO
ISOLATE CAUSE
OF
PROBLEM.