DEMODULATION SETUP CHECKLIST
Here are the steps to set u p demodulation measurements:
1 . Activate demodulation: press MEAS MODE followed by LINEAR RES, press
DEMOD ON OFF to ON, then press DEMOD SELECT. (page 5-6)
2. Select the channel(s) to be demodulated: use the DEMOD SELECT menu.
(page 5-6)
3 . Select AM, FM or PM demodulation: use the demod types menu. (page 5-6)
4. For FM or PM demodulation, select auto carrier or enter a user carrier frequency:
use the PM/FM CRRIER menu. (page 5-8)
5. Set up the delete frequency table (optional): press DELETE FREQ in the DEMOD
SELECT menu. (page 5-9)
6. Proceed with linear resolution measurement setup: refer to Chapter 1 .
7 . Use previewing (optional): press PRVIEW ON OFF to ON. (page 5-10)
8. Use the DEMOD POLAR display, if desired: press MEAS DISP, then FILTRD
INPUT followed by DEMOD POLAR. (page 5-10).