Waters - The Science of What's Possible. No flow from fluidics on Xevo/SYNAPT/Vion - WKB50419 SYMPTOMS • No beam from infusion • With PEEK disconnected from source, no liquid flow is seen when trying to infuse from bottles • Flow can be seen when infusing wash solution (after purging) ENVIRONMENT •...
Page 2
Waters - The Science of What's Possible. CAUSE The orange PEEK tubing does not reach down to the bottom of the 30-mL Nalgene fluidics bottle; therefore, air is drawn instead of sample. FIX or WORKAROUND 1. Open the fluidics door.