Parts Replacement
laçin:
To replace the lamp, the following proce-
dure should be performed:
NOTE: For optimum performance and
extended bulb life, use only Pelton &
Crane bulbs (part no. 017133) identi-
fied by Pelton & Crane on the ceramic
ends.
AWARNING
To avoid burning fingers or hand,
ensure lamp is off and cool.
© Open rear of light by depressing small
pin on side of light frame. Rotate lamp
A 90" so edge can be held with fin-
gers.
© Hold lamp with left hand and press
lever B to right side with right hand.
Remove lamp by gently pulling out.
A CAUTION
Use tissue to hold new lamp while
installing. Do not touch new lamp with
bare fingers. Contamination may
shorten lamp life. If touched, clean
with isopropyl alcohol and dry with
clean cloth to remove fingerprints.
Install replacement lamp with exhaust
tip away from reflector. Close reflector.
Exchanging Autoclavable Handles: To
exchange autoclavable handles C, the
following procedure should be performed:
NOTE: Autoclavable handles are pack-
aged non-sterile. Wash handles thor-
oughly and sterilize before using.
e Remove autoclavable handles by
grasping handles on either side, plac-
ing thumbs underneath slides.
e Retract spring-loaded slides D by
pushing thumbs outward and lift off
handles.
® Grasp spare autoclavable handles C
on either side, placing thumbs under-
neath slides.
® Retract spring-loaded slides D by
pushing thumbs outward.
@ Position handles over frame and
release slide, locking handles into
place.