Siemens PELTON & CRANE Light Fantastic II Installation Instructions Manual page 40

Table of Contents

Advertisement

is
offered
for
dentists
needing
two
intra-oral
lights.
One
light
may
be
placed
over
the
head
and the other in front to overcome difficult light-
ing conditions. Variable length columns are avail-
able to accommodate
8' (2.44 m) to 3' (3.97 m)
ceiling height installations.
1.08
Track Assembly
(Figure 4)-The track as-
sembly for models
LFTII and
LFTI-D
con-
sists
of
a
plywood
pallet,
an
aluminum
track
channel, aluminum dress covers, panels and end
caps.
Proper
installation to ceiling substructures
will ensure
drift free
and
vibration
free operation
of
the
light
head
(see
Installation
Instructions
YL3-096035).
TRACK
TROLLEY
p
|
VERTICAL
COLUMN
\
IES
P
TRANSFORMER
BOX
FIGURE 4 - LFTII
1.09
Transformer
Box
(Figure 4)-Removal
of
the transformer box lid provides access to
the
transformer
for
required
wiring.
A
fuse
lo-
cated
on the end of the transformer
box is pro-
vided for circuit protection of the light. The LFTII
uses
a 3AG
SLOW
BLOW
2 amp
125
volt fuse
(Littlefuse
#313002/S)
and
the
LFTII-D
uses
a
3AG
SLOW
BLOW
4 amp
125 volt fuse (Littlefuse
#313004/S) .
1.11
Vertical
Column
(Figure
4)-The
column
length and
associated
power cable deter-
mine height of light and are the only two variable
factors
with
different
ceiling
heights.
If light
is
moved
to
new
office
with
a
different
ceiling
height,
a new column
and associated
power ca-
ble may
be purchased
to easily modify the light.
FIGURE
5 - LFTII-D
1.12
Column
Stop
(Figure
6)-The
entire
arm
and
light
may
be
rotated
approximately
322° with respect to the column.
Location
of the
column
stop may
be varied to provide
optimum
operating
positions.
For
adjustment
procedure,
refer to the Light Fantastic Installation and Adjust-
ment instructions, YL3-096035.
CAUTION:
DO NOT ROTATE
ARM
MORE
THAN
360°
WITH
RESPECT
TO THE COLUMN
OR TWIST.
ING DAMAGE
TO THE INTER-
NAL WIRING MAY
RESULT.
1.13
Dimmer
Control
(Figure
6)-Rotation
of
the dimmer
control
light intensity adjust-
ment
to be
varied
from
approximately
1500
to
2500 footcancles.
1.14
Refer to page
1, paragraphs
1.02 through
1.06 for familiarization of other important
features of the LFTI| and
LFTII-D.