Remove autoclavable handles by grasp-
ing
handles
on
either
side,
placing
thumbs underneath slides.
e
Retract spring-loaded slides by pushing
thumbs outward and lift off handles.
e
Grasp
spare autoclavable
handles
on
either side, placing thumbs
underneath
slides.
®
Retract spring-loaded slides by pushing
thumbs outward,
e
Position
handles
over
frame
and
re-
lease slide, locking handles
into place.
3.05
Cleaning
and
Sterilizing
Autoclavable
Handles—Handles may be cleaned with a
soft cloth
and
mild
detergent.
Handles
may
be
safely
sterilized
using
the
following
methods;
steam
autoclaving,
dry
heat
sterilization
up
to
320°
F
(160°
C),
ethylene
oxide
sterilization,
gamma
radiation and cold sterilization.
CAUTION:
SOME
COLD
STERILANTS
MAY
DISCOLOR
HANDLES
WITH
USE
OVER
A
PRO-
LONGED
PERIOD OF TIME.
ADJUSTMENT
SCREWS
FIGURE
12 - ADJUSTMENT SCREWS
FOR FOCUSING LAMP
PATTERN
A
PATTERNS PARTIALLY OVERLAPPED AND
ARE NOT FOCUSED FOR OPTIMUM PATTERN.
THIS PATTERN WILL "SHIFT" WHEN
PARTIALLY BLOCKED OUT.
PATTERN
B
PATTERNS ARE SUPERIMPOSED.
FOR "OPTIMUM" PATTERN.
PATTERN WILL NOT SHIFT.
PATTERN
C
IRREGULAR PATTERNS. ONLY
ONE ADJUSTING SCREW
SHOULD BE TURNED.
FIGURE
13 - FOCUSING
LAMP
PATTERNS
Remove
autoclavable
handles
to
ac-
cess adjustment screws.
Turn two screws on front of light clock-
wise
until
both
screws
are
tight.
The
patterns will now
be partially overlapped
as shown
in pattern
A in Figure
13.
Turn
each
screw
counterclockwise
equally one turn at a time until the pat-
terns
are
superimposed
as
shown
in
Pattern B in Figure
13.
Inspect for optimum pattern by blocking
out
one
half of
the
light.
If the
pattern,
appears
to shift,
further
adjustment
is
required.
Otherwise,
focusing
of
the.
lamp
is completed.