With the help of the diagnosis menu, check the function sensor and magnet valve.
Error during Titration
Check that there is enough reagent available.
Check the pipes between the reagent bottle and the dosing-pump for air bubbles. If
necessary, pump reagent until the pipes are full of reagent.
Check that the blue O-ring is still on the nozzle of the dosing-plug.
Check that there is water in the chamber.
Check that there is a magnetic stirrer (stir-fish) in the chamber.
Check the reagent delivery, the sensor and the stir-fish with the diagnosis menu.
False test result
Check that the reagent corresponds to that programmed.
Check for air bubbles in the reagent delivery pipe.
During the blank test, check that the sample water is not coloured by foreign
substances, sediment or air bubbles.
Check that there is water in the chamber and that the supply and waste pipes are not
swopped.
Check that there is a magnetic stirrer (stir-fish) in the chamber.
Check that the blue O-ring is still on the nozzle of the dosing-plug and that it is
positioned correctly.
Check that the magnet valve closes properly.
Exchange the reagent pump-cassette
Re-calibrate the unit with a new correction factor.