Unit description
Analysis Qualities
Measurement method
Measurement range
Accuracy
Reagent consumption
Measurement duration
Number of analyses
Expiration date of the
reagent
Water consumption
Waltron User Manual 101-048-E.1
6051 Hardness Analyzer
Titration method with colour change
Total Hardness
ppm CaCO
0,21 ... 535,7
3
°dH
0,012 ... 30
°f
0,021 ... 53,6
°e
0,015 ... 38,6
mmol/l
0,0021 ... 5,36
The measuring range is defined by the reagent used. The entire
measuring range of the total hardness is achieved by using different
reagents.
Measurement accuracy:
+/- 5% of the upper value of the respective reagent (see page 74.)
Repetition accuracy:
+/- 2.5% of the upper value of the respective.
Please note: The accuracy of the measurements may be adversely affected
by the impurity of the water. In such cases, we recommend that you take a
hardness measurement by hand titration and then calibrate the analysis
unit to this value.
ca. 0.05-0.5 ml/ analysis depending on the hardness of the water
ca. 3 minutes depending on the hardness of the water
ca. 10,000 analyses / 500 ml reagent by low hardness. The consumption is
dependent on the measured water hardness and the reagent used.
At least 2 years
ca. 1l/analysis at 2 bar
The water consumption varies according to pressure and flush-time.
Carbonate Hardness
5,4 ... 401,8
0,3 ... 22,5
0,5 ... 40,2
0,39 ... 28,9
0,11 ... 8,04
13