Configuration
Relay1 (Limit 1)
Period or impulse contact
Your setting:
Define Relay 2
Your setting:
Relay2 (Limit 2)
Period or impulse contact
Your setting:
Define Relay 4
Your setting:
Waltron User Manual 101-048-E.1
6051 Hardness Analyzer
Do you need a period or impulse contact on relay 1?[OK]
You have the following possibilities:
Period / Impulse [OK]
By selecting impulse, you can set the impulse length to
control a switch room:
001 to 999 seconds.
Choose the function of input 2 [OK]
You have the following possibilities:
Limit 2 or reagent level <10%
Via the navigations-keys, make a selection and confirm [OK].
Alternatively, the reagent level can be configured to relay 4.
Do you need a period or impulse contact on relay 2?[OK]
You have the following possibilities:
Period / Impulse [OK]
By selecting impulse, you can set the impulse length to
control a switch room:
001 to 999 seconds.
Choose the function of input 4. [OK]
You have the following possibilities:
Analysis or reagent level <10%
Via the navigations-keys, make a selection and confirm [OK].
With analysis, you can control an external pump, the cool
water valve of a sample cooler or an external controller.
Alternatively, the reagent level can be configured to relay 2.