Configuration
Select reagent
Your setting:
Deliver reagent
Select unit of measurement
Your setting:
Set flushing time
Your setting:
Waltron User Manual 101-048-E.1
6051 Hardness Analyzer
Please select your reagent [OK]
According to test parameters, a choice of possible reagents
is now shown:
Total hardness:
500S/500 to 600/500
Carbonate hardness:
C710/500 to C730/500
Select and confirm with[OK]
Place a new bottle of reagent in position [OK]
The reagent pump starts the delivery of reagent. The
delivery
can
be
stopped
30secondsreagent will have reached the measuring
chamber.
Select the unit in which the test results should be
displayed. [OK] You will get a choice of possible
measurement units according to the measurement
parameters:
°dH, °f, ppm(CACO3), mmol/l, mg/l, mval/l, e
Select and confirm with[OK]
Give in the required flushing time.[OK]
Flushing time: 0001 – 9999seconds
The flushing time must be set according to the length of
the inlet pipe and the water pressure. It must be
sufficiently long to allow a complete change of sample
water in the chamber.
Select with navigation keys and confirm. [OK]