Waltron Aqualert 6051 Instruction Manual page 50

Hardness analyzer
Hide thumbs Also See for Aqualert 6051:
Table of Contents

Advertisement

Test repetition,
First result suppression
Your setting:
Define input2
Your setting:
50
Set the number of tests that you require to be made
following a failed test. [OK]
You have the following possibilities:
No repeat
Repeat once
Repeat twice
Repeat 3 times
Via the navigations-keys, you can select the value and
confirm [OK].
An analysis can produce a bad result after the system has
stood a while. It is, therefore advisable to carry out a control
test to confirm a result before an alarm is set off.
Choose the function of input2 [OK]
You have the following possibilities:
Deactivated, analyse start, Water-meter, Flow-monitor
Via the navigations-keys, make a selection and confirm [OK].
Different functions can be assigned to input2.
Deactivated
If the input is not used
Analysis start
The analysis can be started by a controller. This is only
possible if the unit is ready and not already in the process of
an analysis.
Water meter
This setting should be selected if the unit is to be controlled
by an external water-meter.
Flow-monitor
This setting should be selected if an external flow-monitor is
connected to the input and the analysis unit should only
carry out tests at pre-set time intervals if water is removed
from the system.
Configuration
Waltron User Manual 101-048-E.1
6051 Hardness Analyzer