7.2 Browse the parameter
Long press
for 3 seconds in measurement mode.
Then press or to browse the parameters
1. Working voltage
2. Slope
3. Zero offset
4. Manual Temperature compensation
5. Atmospheric Pressure
6. Salinity
7. Hold Auto Lock
8. Auto power off
9. Battery power
7.3 pH SET UP
7.3.1 P01 Atmospheric Pressure
In measurement mode, press
to switch between Saturation Percentage and Oxygen
concentration mg/L. Press
to enter into parameter setting mode.Press
electrode calibration mode.
In P-01, you can adjust Atmospheric Pressure value for compensation to get the better measure-
ment result.
Please refer to above description steps to setup P-01.
Range: 600mbar-1400mbar.
You can go to next parameter setting by pressing
back to measurement mode.
32
to enter into
, or press
key to quit and go
5.3.2 P02 Manual Temperature setting
After entering P-02, you can set the manual temperature by pressing
for speedy setting. Confirm your setting by pressing
Manual Temperature setting range: 0°C --100°C.
Please refer to above description steps to set up P-02.
You can go to next parameter setting by pressing
back to measurement mode.
Note:
1. Please make sure the Temperature Electrode is UNPLUGGED when manual Temperature
compensation. Otherwise, the unit will display "Err" and go back to the set up menu for fresh
operation. The "Manual" is displayed beneath the main screen normally.
2. It is not working to unplug the Temperature Electrode while setting. You have to set again after
removing the electrode and quit the current mode.
. Long press
.
, or pressing
key to quit and go
09