Technical Specifications; Browse The Parameter; Set Up; P01 Cell Constant (K) - CLEAN Instruments PH200 Operation Manual

Ph/orp/ conductivity/ tds/ salinity/ dissolved oxygen meter
Hide thumbs Also See for PH200:
Table of Contents

Advertisement

6.1 Technical Specifications

Conductivity
Range
0.000
Resolution
0.001
Accuracy
± 0.5% F.S.
TDS
Range
0.000 mg/L 400.0 g/L
Resolution
0.001 mg/L 0.1 g/L
Accuracy
± 0.5% F.S.
Temperature
Temperature
-10.0 °C 110.0 °C
Resolution
0.1 °C
Accuracy
± 0.2 °C
Power
Power supply
2*7 AAA batteries
Working time
>500 hrs
Environment
Humidity
-5 °C 60 °C Relative humidity 90%

6.2 Browse the parameter

Long press
for 3 seconds in measurement mode.
Then press
or
to browse the parameters
1. Cell constant
2. Slope
3. Base temperature
4. Temperature change rate
5. TDS
6. TC Manual Temperature compensation
7. Hold Auto Lock
8. Auto power off
9. Battery power
20
S/cm 400.0 mS/cm
S/cm 0.1 mS/cm

6.3 SET UP

6.3.1 P01 Cell Constant (K)

The unit will enter into measurement mode after power on. Factory default is Conductivity
measurement mode. Press
to switch between Conductivity measurement mode and Total
Dissolved Solids measurement mode. Press
to enter into electrode calibration mode.
In P-01, you can set up cell constant K of your sensor for measurement application.
Please refer to above description steps to setup P-01.
You can go to next parameter setting by pressing
back to measurement mode.
Measurement range:
Cell constant K=1.0, For Middle range;
Cell constant K=10, For Wide range (More than 40 mS or 10ppt)
Cell constant K=0.1, For Middle range (Below 40 s or 10ppm)
Factory default: Cell constant K=1.0.
to enter into parameter setting mode. Press
, or press
key to quit and go