P03 Temperature Offset Setting - CLEAN Instruments PH200 Operation Manual

Ph/orp/ conductivity/ tds/ salinity/ dissolved oxygen meter
Hide thumbs Also See for PH200:
Table of Contents

Advertisement

5.3.3 P03 Temperature Offset setting

After entering P-03, you can set the temperature to the appropriate value by pressing
according to the actual temperature. Long press
setting by pressing
.
Temperature Offset range: +10°C.
Please refer to above description steps to set up P-03.
You can go to next parameter setting by pressing
go back to measurement mode.
Note:
1. Please make sure the Temperature Electrode is PLUGGED when Temperature Offset setting.
Otherwise, the unit will display "Err" and go back to the set up menu for fresh operation. The
"Auto" will be displayed beneath the main screen the screen normally.
2. It is not working to plug the Temperature Electrode while setting. You have to set again after
plugging in the electrode and quit the current mode.
10
for speedy setting. Confirm your
, or pressing
key to quit and
7.1 Technical Specifications
Testing items
Saturation Percentage
Range
0.0% 400.0%
Resolution
0.1%
Accuracy
±0.2% F.S.
Oxygen concentration
0.00 mg/L 40.00 mg/L
Range
Resolution
0.01 mg/L
Accuracy
±0.2% F.S.
Temperature
Range
when compensating:0.0 °C 50.0 °C
when testing:0.0 °C 50.0 °C
Resolution
0.1 °C
Accuracy
±0.2 °C
Settings
Atmospheric Pressure
600 mbar 1400 mbar
Range
1 mbar
Resolution
Default
1013 mbar
Salinity
Range
0.0 g/L 40.0 g/L
Resolution
0.1 g/L
0.0 g/L
Default
Others
Power
Power supply
2*7AAA batteries
Working time
>500 hrs
Environment
-5 °C 60 °C Relative humidity 90%
Humidity
31