5.3 pH Set Up
5.3.1 P01 pH Standard Solution
After power-up, the unit will enter the measurement mode. Factory default is pH measuring
mode. You can switch between the pH measurement mode and the ORP measurement mode
by pressing the
Key.
Press
to enter SET up step; Press
In P-01, you can select a group of buffer solutions as a standard: USA (10.01 7.00 4.01) or NIST
(4.01 6.86 9.18).
Please refer to above description steps to set up P-01.
You can go to next parameter setting by pressing
go back to measurement mode.
Factory default: NIST
08
to enter the Electrode Calibration mode.
, or pressing
key to quit and
7.3.2 P02 Salinity Compensation
In P-02, you can set up Salinity concentration value for compensation to get the better
measurement result.
Please refer to above description steps to setup P-02.
Range: 0.0-40.0mg/L
You can go to next parameter setting by pressing
back to measurement mode.
7.3.3 P03 Temperature Offset
In P-03, you can set up temperature offset value for better measurement result.
Please refer to above description steps to setup P-03.
Range: +10°C
You can go to next parameter setting by pressing
back to measurement mode.
, or press
key to quit and go
, or press
key to quit and go
33