5.3.2 PO2 Manual Temperature setting
*
After entering P-02, you can set the manual temperature by pressing DE
. Long press
DE
for speedy setting. Confirm your setting by pressing ES i
*
Manual Temperature setting range: 0°C-100°C.
*
Please refer to above description steps to set up P-02.
*
You can go to next parameter setting by pressing DE
, Or pressing
key to quit and go
back to measurement mode.
Note:
1. Please make sure the Temperature Electrode is UNPLUGGED when manual Temperature
compensation. Otherwise, the unit will display "Err" and go back to the set up menu for fresh
operation. The "Manual" is displayed beneath the main screen normally.
2. It is not working to unplug the Temperature Electrode while setting. You have to set again after
removing the electrode and quit the current mode.
09
eo
o
LO
25
A.
=
Les
Les
N
ba
Ou
Need help?
Do you have a question about the PH200 and is the answer not in the manual?