CLEAN Instruments PH200 Operation Manual

Ph/orp/ conductivity/ tds/ salinity/ dissolved oxygen meter
Hide thumbs Also See for PH200:

Advertisement

Quick Links

PH200/CON200/DO200
Meter
www.cleaninst.com
Customer Service: CS@cleaninst.com
CLEAN INSTRUMENTS
2006, No.511 Tianmu W. Rd. Shanghai 200070, China
11F No.25,Siwei St., Zhonghe Dist.,New Taipe City 23570, Taiwan
pH/ORP
Conductivity / TDS / Salinity
Dissolved Oxygen
Meter
Operation Manual
MODE
CAL
ESC
INFO
SET
ENT
HOLD
DO200
www.cleaninst.com
MI
MR
2013.07

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PH200 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for CLEAN Instruments PH200

  • Page 1 Conductivity / TDS / Salinity Dissolved Oxygen Meter Operation Manual MODE INFO PH200/CON200/DO200 Meter HOLD DO200 www.cleaninst.com Customer Service: CS@cleaninst.com www.cleaninst.com CLEAN INSTRUMENTS 2006, No.511 Tianmu W. Rd. Shanghai 200070, China 11F No.25,Siwei St., Zhonghe Dist.,New Taipe City 23570, Taiwan...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS 7.7.2 Water saturated dissolved oxygen in different temperature (Atmospheric Pressure 101325Pa 100% RH Salinity 0‰) 1 Preface Temperature°C Dissolved oxygen(mg/L) Temperature°C Dissolved oxygen(mg/L) 1.1 Before Use 14.64 9.08 1.2 In Use 14.22 8.90 1.3 Safety 13.82 8.73 2 Basic Operation 2.1 Battery 13.44...
  • Page 3: Water Saturated Dissolved Oxygen In Different Temperature

    7.7 APPENDIX 6.3.3 P03 Conductivity Temperature Factor 7.7.1 The corresponding Average Atmospheric Pressure with the Altitude 6.3.4 P04 TDS Factor 6.3.5 P05 Temperature modified Factor Altitude h/m Average Atmospheric Pressure ph/kPa Altitude h/m Average Atmospheric Pressure ph/kPa 6.3.6 P06 Auto Lock Function Setting 6.3.7 P07 Auto Power Off Setting 101.3 2800...
  • Page 4: Preface

    1 Preface 7.6.3 Setting parameter 1.1 Before Use Thank you for selecting CLEAN meter. Although the meter use advanced technology and meet the requirements of current safety rules, improper use can still threaten the safety of users, and / or cause harmful influences to factory and other equipments.
  • Page 5: Basic Operation

    The meter will automatically go back to measurement mode after data saved. The unit at most can save 50 sets data. The screen will show “OVER” if more than 50 sets reminds you the PH200 CON200 memory card is full.
  • Page 6: Key Panel

    3 Key Panel 7.5 Measurement Mode 3.1 Key instruction 1. You can first set the meter per your request. Otherwise, the setting will be factory defaults. 2. Before each measuring, first lay aside the electrode for around 15mins to fully activate the Short press: Short press means release the key once after pressing.
  • Page 7: Appearance

    4 Appearance 7.4 Calibration 4.1 Display 1. In measurement mode, press to enter calibration process. . 2. First point calibration-100% air calibration: put the sensor in air, after you get a stable value or the stable symbol shows up, press key to confirm 100.0% saturated dissolved oxygen point.
  • Page 8: Display Character Table

    7.3.7 P07 Reverting to Factory Default Setting 4.2 Display Character Table Sensor Password Setting Automatic Temperature Compensation Factory Defaults Manual Temperature Compensation Sensor Slope 1 Sensor Slope 2 Automatic Temperature Compensation Save Data Manual Temperature Compensation Electrode Slope Error autolock Auto Off no stored data Clear data...
  • Page 9: Ph / Orp

    / ORP 7.3.6 P06 Clearing Memory Function PH200 Meter · New design, Comfortable holding, Easy carrying , Simple Operation · 65*40mm, large LCD with backlight for easy meter information reading · IP67 rated, dustproof and waterproof, Floats on water ·...
  • Page 10: Technical Specifications

    5.1 Technical Specifications 7.3.4 P04 Auto Lock Function Setting Range -2.00 16.00 pH Resolution 0.01 pH Accuracy ±0.2% F.S. Range -2000~2000 mV Resolution 1 mV Accuracy ± 0.2% F.S. Temperature Range -10.0 °C~110.0 °C Resolution 0.1 °C Accuracy ± 0.2 °C Power Power supply 2*7 AAA batteries...
  • Page 11: Ph Set Up

    5.3 pH Set Up 7.3.2 P02 Salinity Compensation 5.3.1 P01 pH Standard Solution In P-02, you can set up Salinity concentration value for compensation to get the better measurement result. Please refer to above description steps to setup P-02. Range: 0.0-40.0mg/L You can go to next parameter setting by pressing , or press key to quit and go...
  • Page 12: P02 Manual Temperature Setting

    7.2 Browse the parameter 5.3.2 P02 Manual Temperature setting Long press for 3 seconds in measurement mode. Then press            or           to browse the parameters 1. Working voltage 2. Slope 3. Zero offset 4. Manual Temperature compensation After entering P-02, you can set the manual temperature by pressing .
  • Page 13: P03 Temperature Offset Setting

    7.1 Technical Specifications 5.3.3 P03 Temperature Offset setting Testing items Saturation Percentage Range 0.0% 400.0% Resolution 0.1% Accuracy ±0.2% F.S. After entering P-03, you can set the temperature to the appropriate value by pressing Oxygen concentration 0.00 mg/L 40.00 mg/L Range according to the actual temperature.
  • Page 14: P04 Auto Lock Function Setting

    Dissolved Oxygen 5.3.4 P04 Auto Lock Function setting DO200 Meter · New design, Comfortable holding, Easy carrying , Simple Operation · 65*40mm, large LCD with backlight for easy meter information reading · IP67 rated, dustproof and waterproof, Floats on water ·...
  • Page 15: P05 Auto Power Off Setting

    6.6 Appendix 5.3.5 P05 Auto Power Off Setting 6.6.1 Conductivity sensor measuring range: Range Cell Constant Resolution 0.000~4.000 S/cm K=0.01 cm 0.001 S/cm 0.00~40.00 S/cm 0.01 S/cm 0.0~400.0 S/cm 0.1 S/cm 0.00~40 .00 S/cm K=0.1 cm 0.01 S/cm 0.0~400.0 S/cm 0.1 S/cm 0~4000 S/cm 1 S/cm...
  • Page 16: P07 Reverting To Factory Default Setting

    6.5 Calibration 5.3.7 P07 Reverting to Factory Default Setting Note: Calibration must be done in the following three occasions: 1. before using new sensors or new meter; 2. when the sensor been used for a long time and lead to a widely deviation measuring result. 3.
  • Page 17: Calibration Mode

    5.4 Calibration Mode 6.4 Setting Parameter 1 pH Calibration change standard solution change standard solution 1. Long press to enter into Setting parameter when in measurement mode. Press 2. Inset the electrode in the pre-set standard solution. You can refer to P-01 explanation on the to check through all the parameters.
  • Page 18: Measurement Mode

    5.5 Measurement Mode 6.3.9 P09 Reverting to Factory Default Setting 1. You can first set the meter per your request. Otherwise, the setting will be factory defaults. 2. Use tap water or distilled water to clean the pH and the temperature electrode, removing the adhesive impurities on the surface of electrode.
  • Page 19: Storage Of Data

    5.6 Storage of Data 6.3.8 P08 Clearing Memory Function 5.6.1. Storage of measuring data 1.If you need to save the measuring data once it is stabilized, press , the Lot no. will be shown on the upper right of the screen. Press to save the data.
  • Page 20: Calibration Parameter And Setting Parameter

    5.7 Calibration Parameter and Setting Parameter 6.3.6 P06 Auto Lock Function Setting 1. Long press to enter into Setting parameter when in measurement mode. Press In P-06, you can set the Auto Lock Function. to check through all the Calibration parameters and Setting parameter. Please refer to above description steps to setup P-06.
  • Page 21: Appendix

    5.8 Appendix 6.3.4 P04 TDS Factor Factory defaults Corresponding interface Item Factory Defaults P-01 Standard Solution USA: 10.01,7.00,4.01 P-02 Temperature Manual Temperature Compensation 25 °C In P-04, you can set TDS factor for your application. P-04 Auto Lock Please refer to above description steps to setup P-04. P-05 Auto power off TDS Factor range: 0.40-1.00...
  • Page 22: Conductivity / Tds / Salinity

    Conductivity / TDS / Salinity 6.3.2 P02 Temperature Compensation Base CON200 Meter · New design, Comfortable holding, Easy carrying , Simple Operation · 65*40mm, large LCD with backlight for easy meter information reading · IP67 rated, dustproof and waterproof, Floats on water ·...
  • Page 23: Technical Specifications

    6.1 Technical Specifications 6.3 SET UP 6.3.1 P01 Cell Constant (K) Conductivity Range 0.000 S/cm 400.0 mS/cm Resolution 0.001 S/cm 0.1 mS/cm Accuracy ± 0.5% F.S. Range 0.000 mg/L 400.0 g/L Resolution 0.001 mg/L 0.1 g/L Accuracy ± 0.5% F.S. Temperature Temperature -10.0 °C 110.0 °C...
  • Page 24 If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return to CLEAN Instruments been the result of abuse or misuse within the warranty period, please return to CLEAN Instruments and amendment will be made without any charge.