P05 Auto Power Off Setting; P06 Clearing Memory Function - CLEAN Instruments PH200 Operation Manual

Ph/orp/ conductivity/ tds/ salinity/ dissolved oxygen meter
Hide thumbs Also See for PH200:
Table of Contents

Advertisement

5.3.5 P05 Auto Power Off Setting

After entering P-05, you can select ON or OFF the Auto Power Off by pressing
Auto Power Off is ON, the meter will power off if no operation in 10mins.
Please refer to above description steps to set up P-05.
You can go to next parameter setting by pressing
back to measurement mode.

5.3.6 P06 Clearing Memory Function

After entering P-06, you can select NO or YES to not clear or clear the data by pressing
If select YES, the meter will clear all the saved data.
Please refer to above description steps to set up P-06.
You can go to next parameter setting by pressing
back to measurement mode.
12
. If the
, or pressing
key to quit and go
.
, or pressing
key to quit and go
6.6 Appendix
6.6.1 Conductivity sensor measuring range:
Range
Cell Constant
0.000~4.000 S/cm
K=0.01 cm
0.00~40.00 S/cm
0.0~400.0 S/cm
0.00~40 .00 S/cm
K=0.1 cm
0.0~400.0 S/cm
0~4000 S/cm
0.0~400.0 S/cm
K=1 cm
0~4000 S/cm
0~40.00 mS/cm
0~4000 S/cm
K=10 cm
0~40.00 mS/cm
0~400.0 mS/cm
6.6.2 Factory Defaults
Instrument Constant
Measurement Range
Cell constant type
0.01, 0.1, 1, 10
Temperature Base
15.0~35.0 °C
Conductivity variance ratio per temperature
0.00~10.00%
TDS factor
0.40~1.00
Salinity factor
0.48~0.65
Temperature bias
+10 °C of the indicated temperature
Auto Lock function
OFF or ON
Auto Power off function
ON or OFF
Resolution
0.001 S/cm
0.01 S/cm
0.1 S/cm
0.01 S/cm
0.1 S/cm
1 S/cm
0.1 S/cm
1 S/cm
0.01 mS/cm
1 S/cm
0.01 mS/cm
0.1 mS/cm
Initial value
1.0
25.0 °C
2.00%
0.50
0.65
0.0 °C
OFF
ON
29

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PH200 and is the answer not in the manual?

This manual is also suitable for:

Con200Do200

Table of Contents