Technical Specifications; Browse The Parameter - CLEAN Instruments PH200 Operation Manual

Ph/orp/ conductivity/ tds/ salinity/ dissolved oxygen meter
Hide thumbs Also See for PH200:
Table of Contents

Advertisement

7.3.4 P04 Auto Lock Function Setting
In P-04, you can set the Auto Lock Function.
Please refer to above description steps to setup P-04.
If the Auto Lock Function is ON, the meter can lock the measured value after the reading had
stabilized and display
. Press
to unlock.
You can go to next parameter setting by pressing
back to measurement mode.
7.3.5 P05 Auto Power Off Setting
In P-05, you can set the Auto Power Off Function.
Please refer to above description steps to setup P-05.
If the Auto Power Off is ON, the meter will power off if no operation in 10mins.
You can go to next parameter setting by pressing
back to measurement mode.
34
, or press
key to quit and go
, or press
key to quit and go

5.1 Technical Specifications

pH
Range
-2.00
16.00 pH
Resolution
0.01 pH
Accuracy
±0.2% F.S.
ORP
Range
-2000~2000 mV
Resolution
1 mV
Accuracy
± 0.2% F.S.
Temperature
Range
-10.0 °C~110.0 °C
Resolution
0.1 °C
Accuracy
± 0.2 °C
Power
Power supply
2*7 AAA batteries
Working time
>500 hrs
pH Buffer Kinds
NIST
4.00
6.86
USA
4.01
7.00
Others
Environment
-5 °C~60 °C, Relative humidity 90%
Memory
50 sets

5.2 Browse the parameter

Long press
for 3 seconds in measurement mode.
Then press            or           to browse the parameters
1. Zero offset
2. Slope
3. High/Low segment of the slope
4. Buffer kinds
5. Temperature compensation
6. Hold Auto Lock
7. Auto power off
8. Battery power
9.18
10.01
07

Advertisement

Table of Contents
loading