Quick Start Guide
Note
If the L_TYPE selected in
Step 5
steps will automatically be skipped.
2. To select the signal conditioning L_TYPE from the drop-down
menu:
3. To set XD_SCALE to the 0% and 100% scale points (the
transmitter range):
If L_TYPE is Direct, the AI Block may be placed in mode
to return the device to service. Guided setup does this
automatically.
4. If L_TYPE is Indirect or Indirect Square Root, set OUT_SCALE to
change engineering units.
If L_TYPE is Indirect, the AI Block may be placed in AUTO
mode to return the device to service. Guided setup does this
automatically.
24
are not needed. If guided setup is used, any unneeded
Select Direct for pressure measurements using the device
default units.
Select Indirect for other pressure or level units.
Select Indirect Square Root for flow units.
a) Select the XD_SCALE_UNITS from the drop-down list.
b) Enter the XD_SCALE 0% point.
This may be elevated or suppressed for level
applications.
c) Enter the XD_SCALE 100% point.
This may be elevated or suppressed for level
applications.
a) Select the OUT_SCALE UNITS from the drop-down list.
b) Set the OUT_SCALE low value.
This may be elevated or suppressed for level
applications.
c) Set the OUT_SCALE high value.
This may be elevated or suppressed for level
applications.
Step 2
is Direct, Step 3, Step 4, and