• To prevent the uncontrolled flow of medication, use a CADD
Anti-Siphon Valve, a CADD
®
or a CADD
Administration Set with an attached Add O n Anti-Siphon Valve.
• Use of a syringe with the CADD
DELIVERY of medication. Syringe function can be adversely affected by variations
in plunger dimension and lubricity, which can result in greater force being r equired
to move the plunger. A syringe will lose plunger lubrication as it ages and, as a
result, the amount of under-deliver y will increase which could, on occasion, be
significant. Therefore, the type of medication therapy and delivery accuracy required
must be considered when using a syringe with the CADD
Clinicians must regularly compare the volume remaining in the syringe to the
pump's displayed values such as RES VO L and GIVEN in order to determine
whether under-deliver y of medication is occurring and, if necessar y, take appropriate
action.
2.1.2
CAUTIONS
• This device is not intended to be used for delivery of blood or cellular blood
products.
• This device may interfere with ECG equipment. M onitor ECG equipment carefully
when using this device.
• The pump is not sterile. It is not designed to be sterilized. Sterilization could
damage the microcomputer and other pump parts.
• The pump should be routinely cleaned and kept free of dirt, liquids, and foreign
objects.
• Do not store the pump at temperatures below -40°C (-40°F ) or above 55°C
(131°F).
• Do not operate the pump at temperatures below +2°C (35°F) or above 40°C
(104°F).
• Do not expose the pump to humidity levels above 90% R. H .
• The pump is water resistant. H owever, total immersion is not recommended be-
cause moisture buildup within the case may damage the parts. Do not use pump
in the shower, sauna, or steam bath.
• Do not store the pump for prolonged periods with a battery; the battery could
leak and damage the pump.
• Avoid using the pump in close proximity to sources of strong static electricity or
strong electromagnetic fields.
• The use of a Deltec Pump Pouch is recommended. If the pump is dropped or
inadvertently hit against a hard surface, the pump pouch is designed to minimize
the need for servicing.
®
Administration Set with integrated anti-siphon valve,
®
Administration Set may result in UN DER-
®
Extension Set with
®
pump.