Setting The Continuous Rate (Ml) - SIMS Deltec CADD-1 5100 HFX Operator's Manual

Ambulatory infusion pump
Table of Contents

Advertisement

3.3.2 Setting the Continuous Rate (ML)

The delivery rate in ML (milliliters in a 24-hour period) is the constant rate at
which the pump delivers medication. The delivery rate range is from 001 ml/24 hrs
to 299 ml/24 hrs in 1 ml increments/24 hrs.
Press and release the SELECT MODE key until RATE and ML
appear on the display with a number within the range 001 to
299 ml.
Press the SCROLL keys to the number of milliliters of medication
you wish to deliver in a 24-hour period.
Press and release the SET/CLEAR key within 15 seconds to set
the rate.
NOTE:
If you do not press and release the SET /CL EA R k ey w ithin 15 seconds, the
pum p's m em ory w ill retain the previous delivery rate.
18
RATE
018
*
STOP
ML
´
Î
¤

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CADD-1 5100 HFX and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents