Programming The Cadd-1 Pump For High Flow Delivery - SIMS Deltec CADD-1 5100 HFX Operator's Manual

Ambulatory infusion pump
Table of Contents

Advertisement

3.4
Programming the CADD-1
You can select High Flow Mode so the pump delivers medication at a rate of
approximately 90 ml/hr. Once you select the High Flow Mode and start the
pump, you cannot change the Lock Level or view any other values. High Flow
delivery will continue until you stop the pump or until the reservoir-residual
volume (RES VOL) reaches zero.
In order to program the pump for high flow delivery, follow this procedure:
Confirm that the pump is in the Stop mode and in Lock Level 0. Set the reservoir-
residual volume as described in Section 3.3.1. Attach the cassette and prime
the fluid path as described in Sections 3.5 and 3.6.
Press and hold the PRIME key until three P's appear on the
display. Release the PRIME key.
Press and hold the SCROLL key until
Press and release the PRIME key. "HI F" will appear on the
display.
Press and release the STOP/START key to
start the delivery of medication.
NOTE:
If any one of the above steps is not com pleted w ithin 15 seconds,
the pum p w ill revert to the previous setting.
20
®
Pump for High Flow Delivery