4.2
Pump Maintenance and Cleaning
CAUTION:
• Do not immerse the pump in cleaning fluid or w ater. Do not allow solution to
soak into the pump, accumulate on the keypad, or enter the battery compartment.
• Do not clean the pump w ith acetone, other plastic solvents, or abrasive cleaners.
Use any of the following solutions to clean the pump and accessories:
Soap solution
•
Benzalkonium Chloride concentrate (0.13%)
•
Glutaral Concentrate, USP (2%)
•
10 percent solution of household bleach (one part household bleach to
•
nine parts water)
Alcohol, USP (93%)
•
Isopropyl Alcohol, USP (99%)
•
1. Dampen a soft, lint-free cloth with cleaning solution. Apply the solution to
exterior surface of the pump. Do not allow the solution to soak into the
pump.
2. Wipe the entire surface dry with another soft, lint-free cloth. Allow the pump
to dry completely before use.
4.3
Equipment Exposure to Radiation or Magnetic
Resonance Imaging (MRI)
The CADD-1
®
infusion pump, Model 5100 HFX, is not affected by exposure to
diagnostic radiographic and fluoroscopic radiation.
CAUTION:
The pump SHO ULD NO T BE DIRECTLY IRRADIATED by therapeutic levels
of ionizing radiation because of the risk of permanent damage to the pump's
electronic circuitry. The best procedure to follow is to remove the pump from
the patient during therapeutic radiation sessions. If the pump must remain in
the vicinity during a therapy session, it should be shielded, and its ability to
function properly should be confirmed follow ing treatment.
Magnetic fields produced by magnetic resonance imaging (MRI) equipment
may adversely affect the operation of the pump. Remove the pump from the
patient during MRI procedures and keep it at a safe distance from magnetic
energy.