STEP 6: Begin operation of the current program by pressing and holding the
STOP/ START key to enter the Start mode (Section 3.8), or proceed to
Section 3.2 to program the pump.
NOTE:
You w ill not be able to change the delivery rate until the infusion period has been
com pleted, unless you stop the pum p and reprogram in L ock L evel 0.
The battery's life is dependent on the amount of medication delivered and the
temperature. At the infusion rate of one 50 ml M edication Cassette
per day, an alkaline battery will usually last about seven days. If you use an
®
ULTRALIFE
lithium battery, you will have power for approximately ten days.
Be sure to stop the pump before removing the battery, or up to 10 ml of solution
may not be accounted for by the ML GIVEN function. A battery's power will be
quickly depleted at temperatures below +10°C (50°F).
CAUTION:
Do not store the pump for prolonged periods of time w ith the battery installed.
Battery leakage could damage the pump.