AC Line Power Supply Requirements
AC Line Power Supply Requirements - Overview
Line Power Requirements
Line Power Cable
Keysight N77xx Series
This secton provides information on:
•
Line Power Requirements
•
Line Power Cable
•
Changing the Fuse
The instrument complies with installation category II and can operate
from the single-phase AC power source that supplies between 100 V and
240 V at a frequency in the range 50 to 60 Hz. The maximum voltage
fluctuation is 10% of the nominal supply voltage. The maximum power
consumption is 60 VA with all options installed.
In accordance with international safety standards, the instrument has
a three-wire power cable. When connected to an appropriate AC power
receptacle, this cable earths the instrument cabinet. For the part
number of the power cable for your country and instrument, see
"Accessories - Overview" on page
Getting Started
The recycling symbol indicates the
general ease with which the instrument
can be recycled.
Do not dispose in domestic household
waste. To return unwanted products,
contact your local Keysight office, or
see
http://about.keysight.com/en/
companyinfo/environment/
takeback.shtml
for more information.
69.
1
15