Humidity; Altitude; Pollution Protection; Instrument Cooling - Keysight N77 Series N7711A User Manual

Single-port tunable laser system source
Table of Contents

Advertisement

1
Getting Started

Humidity

Altitude

Pollution Protection

Instrument Cooling

18
The operating humidity is 15 to 95%, non-condensing.
The maximum operating altitude is 2000 m.
The instruments are designed for pollution degree 2.
The instrument has a cooling fan mounted internally.
Mount or position your instrument upright and horizontally, as shown
in Figure 4 so that air can circulate through it freely.
Operating Position
When operating the instrument choose a location that provides at least
75 mm (3 inches) of clearance at the rear, and at least 25 mm (1 inch)
of clearance at each side. Failure to provide adequate air clearance may
result in excessive internal temperature, reducing instrument
reliability. The instrument should not be operated when resting on its
rear or side panels.
Keysight N77xx Series

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7714aN7721aN7722a

Table of Contents