Internal Calibration/Adjustment
Internal
First set either
cal./adjustment
Selection mode
(factory
setting) in the Setup menu
(Balance/scale functions:
Calibration/adjustment: CAL key
function).
Inside the balance housing are built-in
motorized calibration/adjustment
weights.
The internal calibration/adjustment
sequence is as follows:
Unload the balance
Select the calibration function: Press
Cal
the
soft key and then the
Start
soft key
> The internal calibration weight
is applied automatically
> The balance is calibrated
Calibrate,
> If the setting
then auto adjust
setting) is selected in the Setup menu,
the balance is now automatically
adjusted
Calibrate,
> If the setting
then manual adjust
is selected in the Setup menu,
the internal calibration routine is now
ended without adjusting the balance
(otherwise, see "Calibration and
Adjustment Sequence," on next page)
> The internal calibration weight
is removed
> ISO/GLP printout:
see page 131
52
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Internal Linearization
In the Setup menu (Balance/scale func-
or
tions: Calibration/adjustment: CAL key
function), you need to have set
Internal linearization
or select this using the
mode
(factory setting).
Inside the balance housing are built-in
motorized calibration/adjustment
weights for linearization.
The internal linearization sequence
is as follows:
Unload the balance
Select the linearization function: Press
Cal
the
key, and then the
> The internal linearization weights are
applied automatically.
> The balance is linearized
(factory
> The internal linearization weights are
removed
> The balance is adjusted automatically
following internal linearization
> ISO/GLP printout: see page 131
Selection
Select
soft key, the
soft
Start
soft key