ME215/235/414/415/614 Models:
N.I.C.E. Static Electricity Eliminator
(Ionizer) i
Purpose
Ionization of air (i.e., charging the air
with ions of both positive and negative
charges) causes the air surrounding a
sample to become electrically conductive.
Charges are neutralized in the air stream
or grounded.
Be especially careful when weighing
electrostatically charged samples.
Features
– The ionizer can be turned on and off
independently of the position of the
draft shield doors
– When the draft shield doors are open,
the static electricity eliminator (ionizer)
will shut off after the preset time has
elapsed. However, the ionizing time will
start over if the static electricity
eliminator is still running when a door
closes.
– The static electricity eliminator (ionizer)
can be configured in the Setup menu so
that it will stay on only for a certain
time.
– When the ionizer is active, this is
indicated by the flashing symbol J
(flashes from the outside toward the
inside and vice versa)
Factory Setting for the Parameters

Static electricity eliminator (ionizer):

On; Auto off
after 10 sec.
Configuring the Static Electricity Elimina-
tor (Ionizer)
Ionizer
Off
On
Auto-off time:
1 to 99 seconds
Function key
i
Ionizer key
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Password
You can enter a password to block access
to the menu parameter settings and to ID
code and exact calibration weight inputs.
See the detailed description in the chapter
on "Configuring the Balance: Setting the
Device Parameters."
User ID
You can enter your own personal password
(20 characters max.)
Clock
ISO/GLP printouts in particular must be
generated with the date and time stamp
of the specific measurement. This date and
time stamp is optional on other printouts.
See the chapter on "Configuring the
Balance: Setting the Device Parameters."
Interfaces
Purpose
This item enables you to set the
parameters for the following interfaces:
– Serial communications port
– Serial printer port
– External switch function
– Control port function
Serial Communications Port
You can set the serial communications
port to use for the following modes:
SBI
XBPI
Sartonet
Serial Printer Port
You can set the serial printer port to use
for the following printers:
YDPO1IS
YDP02
YDP03
YDPO1IS Label
YDPO2IS
YDP02IS Label
Universal
YDP04IS
YDP04IS Label
External Universal Remote Switch
You can connect an external universal
remote switch (foot switch or bar code
scanner or an external keyboard) to one
of the two serial ports. Then you can
assign one of the following functions to
be performed when
the switch is activated:
Print key
Tare key
Cal key
F1 function key
CF key
F2 function key
Bar code scanner/extra
keyboard
(requires a special
connecting cable)
Ionizer key
Right draft shield key
Left draft shield key
45