Define the Calibration Weight
Step
1. Select Setup menu
2. Select "Balance/scale functions"
3. Select "Calibration/adjustment"
4. Select parameter for external
weight
5. Select "Cal./adj. wt."
6. Enter calibration weight
(e.g., 200.00000 g)
and save
7. Save the calibration weight
8. Exit the Setup menu
54
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Press key(s) (or follow instructions)
s
soft key
O
soft key
O
soft key 5 x
q
soft key
O
soft key
q
2 0 0 . 0
0 0 0 0
soft key
l
soft key
oo
Display/Output
100.00000 = last setting selected
Max 210 g
0%
000000
S
Cal
d=0.01mg
100%
g