Error Messages - Xylem wtw CarboVis 701 IQ TS Operating Manual

Optical sensor for carbon sum parameters and total suspended solids (tss)
Table of Contents

Advertisement

CarboVis 70x IQ TS
Message code
EA2541
EA3541
EAF541
EAI541
EAM3Ax
EC33Ax
EI1541
EI2541
ES2541
ES3541
ba75950e03
05/2017

9.1.1 Error messages

Message text
Sensor temperature too high!
* Check process and application
Sensor temperature too low!
* Check process and application
Optical measuring range exceeded
* Check process (TSS or measured value too high)
* Select measuring location free of air bubbles
* Remove foreign matter from measuring gap
* Clean sensor
* Optimize cleaning settings
* Switch on signal smoothing, increase response time
Zero adjustment erroneous
* Clean the sensor and sensor sleeve repeatedly
* Position the sensor sleeve according to operating manual
* Rinse the measuring chamber with ultrapure water several times
* Fill the measuring chamber with ultrapure water
* Repeat the zero adjustment
Meas. range exceeded or undercut
* Check process
User calibration erroneous,
Check raw value / reference value pairs:
* Set all values within the measuring range
(see operating manual)
* Enter value pairs in ascending order
Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components, replace components
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections
* Defective components,
replace components
Humidity to high in sensor
* Contact service
Failure optical measurement: deviation in reference channel too high
* Execute zero ajustment
Indexes
53