Commissioning
Vertical installation
Protection against
shocks
12
For vertical installation, mount the sensor in a position that allows interfering ele-
ments in the measuring gap to escape with the current. In the case of vertical
installation, the minimum flow rate is 0.2 m/s so the measuring gap is kept clean
optimally.
Measuring gap opening
Fig. 3-2
Vertical installation
When installing the sensor, always make sure it cannot knock against a wall or
another obstacle. Make sure the safety margins are sufficient (e. g. with sus-
pended mounting).
Mount the enclosed shock protectors (see section 3.3.2). Their soft rubber com-
pound absorbs hard impacts.
Fig. 3-3
Sensor with mounted shock protectors