Overview FDO 4410(W) Overview Structure and function FDO 4410 Structure FDO 4410W Shaft Connecting head with active sensor electronics Connection cable (FDO 4410) IDS plug (FDO 4410W) Protective hood Sensor membrane SC-FDO 4410 sensor cap with memory chip Gold-plated contacts for the memory chip of the sensor cap Locking device Measurement window O-ring...
FDO 4410(W) Overview Sensor cap with A memory chip is integrated in the sensor cap. The following data are stored memory chip on the memory chip: Type designation of the sensor cap Series number Data of the factory calibration or user calibration Intelligent For each membrane, the individual calibration values are determined by membrane (IQMC...
Commissioning, measuring FDO 4410(W) Commissioning, measuring General information on the handling of the sensor cap Despite its exterior robustness, the sensor is a high precision optical instru- ment. Therefore, you should take the following precautions when dealing with the FDO 4410(W): ...
FDO 4410(W) Commissioning, measuring Opening and closing the IDS plug connection (FDO 4410W) This section only applies to the IDS plug variant, FDO 4410W Plug connection closed and locked Guidance Groove Opening the plug If necessary, clean the plug connection connection ...
Commissioning, measuring FDO 4410(W) Measuring Minimum depth of Observe the required minimum immersion depth (see chapter 5 T ECHNICAL immersion DATA Incident flow The FDO 4410(W) D.O. sensor enables precise measurements without any incident flow. However, an incident flow of the sensor membrane improves the responding behavior of the sensor (see chapter 5 T ).
FDO 4410(W) Commissioning, measuring the lid. After inserting the sensor, wait for the temperatures of the sensor and cali- bration vessel to adjust. The steps of the check or user calibration are described in detail in the operating manual of the meter. Storage Always store the sensor in the check and storage vessel (FDO Check) at a temperature in the range 0 ...
Maintenance, cleaning, replacement FDO 4410(W) Maintenance, cleaning, replacement General maintenance instructions Handling of the Despite its exterior robustness, the sensor is a high precision optical instru- sensor cap ment. Therefore, special care should be taken when doing any maintenance or cleaning work: ...
FDO 4410(W) Maintenance, cleaning, replacement Exchanging the sensor cap Before removing the sensor cap, observe the general mainte- nance instructions in section 3.1. To exchange the sensor cap, proceed as follows: Removing the Pull the sensor out of the sample. sensor cap Clean the outside of the sensor (see section 3.3.1).
Maintenance, cleaning, replacement FDO 4410(W) Cleaning the sensor 3.3.1 Exterior cleaning Dirt on the sensor can affect the measuring characteristics. Biological depos- its for example, consume oxygen and can, when occurring on the sensor membrane, impair the responding behavior and cause values that are too low.
FDO 4410(W) Maintenance, cleaning, replacement If the sensor cap is visibly damaged it has to be replaced. Checking the zero point of the sensor There are two methods to check the zero point of the sensor: Measurement in a nitrogen atmosphere (recommended method) ...
What to do if... FDO 4410(W) What to do if... Error symptom Cause Remedy No temperature value – No connection between meter – Establish connection between or D.O. value and D.O. sensor meter and D.O. sensor – Cable defective – Return the D.O. sensor Measured value too –...
FDO 4410(W) Technical data Technical data General features Measuring principle Optical measurement based on photolumi- nescence. Temperature sensor Integrated NTC 30 (30 kat 25 °C / 77 °F) Dimensions FDO 4410 (in mm) 205.5 150.0 15.3 21.7 FDO 4410W 252.1 150.0 15.3 21.7...
Page 16
Technical data FDO 4410(W) Connection cable Lengths FDO 4410: 1,5 / 3 m (FDO 4410) FDO 4410W: 1,5 / 3 m Diameter 4.3 mm Smallest allowed Fixed installation: 20 mm bend radius Flexible use: 60 mm Plug type Socket, 4 pins IDS plug Connection type 4-pole, watertight plug connection with lock,...
Page 17
FDO 4410(W) Technical data Characteristic data Accuracy of D.O. measure- ± 1.5 % on delivery ment at 20 °C (68 °F) in air- saturated water Zero signal ≤ 0.02 mg/l O ≤ 0.2 % O saturation ≤ 0.4 mbar O partial pressure Response time at 20 °C (90 % of the final value display after)
Wear parts and accessories FDO 4410(W) Wear parts and accessories Wear parts and Description Model Order no. maintenance SC-FDO 4410 Replacement sensor cap 201 310Y equipment Connection cable Description Model Order no. FDO 4410(W) IDS connection cable, 1.5 m IDS-CABLE- 903 850Y - meter IDS connection cable, 3 m...
FDO 4410 Contact Information Contact Information Ordering & Technical Support Telephone: (800) 897-4151 (937) 767-7241 Monday through Friday, 8:00 AM to 5:00 PM ET Fax: (937) 767-1058 Email: environmental@ysi.com Mail: YSI Incorporated 1725 Brannum Lane Yellow Springs, OH 45387 Internet: www.ysi.com When placing an order please have the following information available: YSI account number (if available)
Page 20
Contact Information FDO 4410 ba76197d01 10/2017...
Page 22
150 countries, we have strong, long-standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise, backed by a legacy of innovation. For more information on how Xylem can help you, go to www.xyleminc.com 1725 Brannum Lane Yellow Springs, OH 45387 Tel: +1 937-767-7241;...