Message code
IA23Ax
IAC541
IC3541
IC4541
IC5541
IC6541
ba75950e03
05/2017

9.1.2 Informative messages

Message text
Calculation of negative concentrations
* Perform user calibration or check user calibration
Zero adjustment was successfully carried out
* Carry out sensor check
* If required, carry out new user calibration
Factory calibration has been activated. Make sure the sensor operates
correctly.
The last valid user zero adjustment was activated. Make sure the sen
sor operates correctly.
Invalid user zero adjustment was replaced by last valid user calibration.
Caution! Wrong measured values possible. Carry out a new successful
zero adjustment to make sure the sensor operates correctly.
The invalid user zero adjustment was replaced by the factory zero
adjustment. Caution! Wrong measured values possible. Check
whether the sensor operates correctly or carry out a new successful
user zero adjustment.