Model 5383 PFPD Quick Reference Guide
Setting Flows for the PFPD
1. Turn off the detector base heater, all
detector gases, and the detector controller.
2. Allow the detector to cool and close the
needle valve.
3. Install the barbed flow adapter in place of
the flame arrestor.
3. Measure the column flow. It should be
<2 mL/minute for optimum performance.
4. Open needle valve four turns.
Measure flows with a flow meter. DO NOT
use the set point on the EPC.
5. Add 11.5 mL/minute H
. Be sure to add the
2
column flow.
6. Add 10 mL/minute of Air 1.
7. Add 10 mL/minute of Air 2 (makeup).
8. Re-install the flame arrestor and turn on the
detector heater. Set it to 250 °C for sulfur
or 300 °C for phosphorus.
9. When the detector temperature reaches
100 °–150 °C, turn on the controller. Pulsing
should start within a few seconds.
10. Record the initial settings.
Have any questions? Call us and let us
help! (800) 336-1911
Suggested Board Parameters
PMT voltage
Set PMT Voltage to 500 V. Monitor
emission, a d j u s t P M T v o l t a g e
s e tt i n g t o ad j us t un ti l t h e f i rst
sharp emission (at about 0.5 ms)
t o ~ 9 0 0 , 0 0 0 c o u n t s . T h e P M T
voltage can vary depending on the
application.
Igniter current 2.8
Trigger level
75000 (initially), adjust to mid span.
For most sulfur applications:
Parameter
Channel 1
Mode
Sulfur
Interpolation
Spline
Multiplier
32
Output zero
<100*
*See the Model 5383 PFPD Operator's Manual Appendix
B for instructions on how to properly set the zero offset
and multiplier for each channel.
1. Confirm installation of the blue, BG-12 filter or the clear, UV-34 filter.
2. Introduce a constant sulfur source for tuning (e.g., a permeation device, PN 290411).
OI Analytical recommends tuning the gas flows using a sulfur emission, regardless of
the application.
3. After the sulfur equilibrates and the sulfur emission is on scale, click Hold in PulseView
Monitor.
4. Adjust H
flow up or down with approximately 1 mL/minute adjustments until the
2
sulfur emission extends to 25 msec (if using a 3-mm combustor, the emission only
extends to 18–20 msec). See the Model 5383 PFPD Operator's Manual for more
information.
5. Use Air 2 (makeup) to adjust frequency to between 2.5-4.0 Hz. Re-adjust H
as necessary.
6. Slowly close the needle valve by turning it clockwise until the PFPD goes into tick-tock
mode.
7. Open the needle valve by turning it counterclockwise until the PFPD is out of tick-tock
mode, then open another ½- to ¾-turn. Attaining tick-tock mode is sometimes not
possible. In that case, just open the needle valve four turns.
8. Record the final settings.
9. For phosphorus applications, install the yellow RG-495 filter and increase Air 1 by
1 mL/minute.
10. Adjust the gate and board parameters for the chosen application.
* These settings are suggested starting points for adjusting the gates. Adjust-
ing the gate start time to eliminate the background HC emission may be
necessary. This is especially important for phosphorus applications. Use
the dual gate subtraction feature for matrices with high sulfur backgrounds
Channel 2
(e.g., garlic, onion, etc.). See Chapter 5 in the Model 5383 PFPD Operator's
Carbon
Manual and application note 1127 at www.oico.com for details.
Spline
For more information contact OI Analytical Technical Support at
1
(800) 336-1911 for assistance.
<10,000*
Tuning the PFPD
Suggested Gate Parameters
Element
Gate
Sulfur
6–24 msec
Carbon
1–3 msec
Phosphorus*
4–15 msec
/Air 1 ratio
2
Square Root
On
Off
Off

Advertisement

loading

Summary of Contents for Xylem OI Analytical 5383

  • Page 1 Model 5383 PFPD Quick Reference Guide Setting Flows for the PFPD Tuning the PFPD 1. Turn off the detector base heater, all 1. Confirm installation of the blue, BG-12 filter or the clear, UV-34 filter. detector gases, and the detector controller. 2.
  • Page 2 Model 5383 PFPD Assembly PMT cap Ignitor assembly Flame arrestor (282624) (286393) PMT signal cable Detector cap (246900) Detector cap mounting screws PMT housing (295287C) (286096) Detector body Combustor PMT HV cable (301333) (282939C, 2 mm or 282913C, 3 mm) Detector body mounting screws (295295C)