Measurement / Operation
Menu item
Cal - # raw value 1
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Calibration
evaluation
30
The entry of calibration values outside the measuring range leads
to a calibration error.
Default values are marked in bold.
Settings
CarboVis:
-10000,00 ... 0,00 ... 20000,00
SolidVis:
-10000,00 ... 0,00 ... 5000,00
CarboVis:
0,00 ... 20000,00
SolidVis:
0,00 ... 5000,00
CarboVis:
-10000,00 ... 0,00 ... 20000,00
SolidVis:
-10000,00 ... 0,00 ... 5000,00
CarboVis:
0,00 ... 20000,00
SolidVis:
0,00 ... 5000,00
The entered calibration data are checked for plausibility by the system. The
check can have the following results:
Display
Meaning/actions
Measured
Successful new valid calibration value pairs were entered for
value display
the sensor.
"----"
Calibration value pairs not accepted. Sensor is blocked for
measurement. A log book message is generated.
Causes:
 Ascending order not observed.
 At least one value is outside the measuring range.
Explanations
Raw value of the first value pair
(lower concentration).
For single-point calibration, enter 0.
Reference value of the first value pair
(lower concentration).
For single-point calibration, enter 0.
Raw value of the second value pair
(higher concentration) or raw value of the
single-point calibration.
Reference value of the second value pair
(higher concentration) or reference value
of the single-point calibration.
CarboVis 70x IQ TS
ba75950e03
05/2017